Water Quality Meter
What is the EC measuring range for 86021?
What can I do if the pH electrode dried out?
Welcome to download the details from the following linkage:
How to clean pH electrode if it's dirty?
Welcome to download the details from the following linkage:
What does a TDS reading actually tell me about my water?
Answer: A TDS reading tells you how much dissolved ionic material is in your water, estimated from its electrical conductivity; it does not tell you which substances they are or whether the water is safe to drink. In my 20 years of field testing, the most common mistake I see is comparing TDS numbers from two meters without checking the conversion factor. A TDS meter measures EC and multiplies it by a factor, so the same sample can read 500 ppm on one meter and 700 ppm on another. A rising TDS is a useful early warning that something has changed, such as salt building up in a fish pond or an RO membrane wearing out, but identifying specific contaminants needs a lab test. For field work I use the AZ 8699 6-in-1 pH EC Salt SG Pen. Its TDS factor is adjustable from 0.40 to 1.00, so I can match the factor my lab or customer uses. It also measures EC, pH, salinity, specific gravity and temperature, with automatic temperature compensation from 0 to 50 °C normalized to 25 °C, and it is rated IP65. Always record the TDS factor together with the reading. Product page: https://www.az-instrument.com.tw/en/product-728950/6-in-1-pH-EC-Salt-SG-Pen-8699-AZ-EB.html
What pH meters can do in daily life?
Many friends ask me what pH meter can help to improve daily life quality.
That is a good question so I collect many documents and summarize them into one simple chart as below.
Hope this is also helpful for you. If you need pH meters, you may check some options from AZ Instrument Corp
Welcome to download the more details from the following linkage,
That is a good question so I collect many documents and summarize them into one simple chart as below.
Hope this is also helpful for you. If you need pH meters, you may check some options from AZ Instrument Corp
Welcome to download the more details from the following linkage,
What kind of PH probes AZ could provide?
For the Benchtop Meter, Portable pH Meter, AZ could provide the following accurate pH probe,
1. 86P3: ATC pH probe
2. 86P2: pH probe without ATC function
3. 86P4: pH probe for Low Ion Strength without ATC function
4. 86P6: Full glass pH probe without ATC function
5. 86P7: Sharp tip pH probe without ATC function
6. 86P1: Flat surface pH probe without ATC function
7. 86P11: 6mm glass pH probe without ATC function
For the AZ Combo Water Testing Meters (ex: 86031, 86021, 84051), their pH probe is "863P".
Welcome to download the more details from the following linkage,
1. 86P3: ATC pH probe
2. 86P2: pH probe without ATC function
3. 86P4: pH probe for Low Ion Strength without ATC function
4. 86P6: Full glass pH probe without ATC function
5. 86P7: Sharp tip pH probe without ATC function
6. 86P1: Flat surface pH probe without ATC function
7. 86P11: 6mm glass pH probe without ATC function
For the AZ Combo Water Testing Meters (ex: 86031, 86021, 84051), their pH probe is "863P".
Welcome to download the more details from the following linkage,
D.O. probe troubleshooting
1) Normal temp. but E03/E02 on D.O. Reason:
a) Low on electrolyte
Solution: See page 16 of operation manual to refill the electrolyte and run 100% calibration.
b) Membrane defect
Solution: See membrane change procedure and run 100% calibration.
c) D.O. sensor damage
Solution: Could not be fixed by user
2) E01 on temp. and D.O. reading Reason:
a) The probe is disconnected
Solution: Plug on the probe and make sure a good connect
b) The wire inside the probe connector is broken.
Solution: Could not be fixed by user.
c) The inside wire of the probe is broken
Solution: Could not be fixed by user.
3) E03 on temp . and E04 on D .O. reading Reason:
a) The temp. sensor is damaged.
Solution: Could not be fixed by user.
b) The temp. related wire inside the probe connector is broken.
Solution: Could not be fixed by user.
4) E17 on D.O. Reading Reason:
a)100% saturation calibration error
Solution: see membrane change procedure and run calibration again.
5) E31 on temp. and D.O. Reading Reason:
a)The inside wire of the probe is broken
Solution: Could not be fixed by user.
Welcome to download the more details from the following linkage,
a) Low on electrolyte
Solution: See page 16 of operation manual to refill the electrolyte and run 100% calibration.
b) Membrane defect
Solution: See membrane change procedure and run 100% calibration.
c) D.O. sensor damage
Solution: Could not be fixed by user
2) E01 on temp. and D.O. reading Reason:
a) The probe is disconnected
Solution: Plug on the probe and make sure a good connect
b) The wire inside the probe connector is broken.
Solution: Could not be fixed by user.
c) The inside wire of the probe is broken
Solution: Could not be fixed by user.
3) E03 on temp . and E04 on D .O. reading Reason:
a) The temp. sensor is damaged.
Solution: Could not be fixed by user.
b) The temp. related wire inside the probe connector is broken.
Solution: Could not be fixed by user.
4) E17 on D.O. Reading Reason:
a)100% saturation calibration error
Solution: see membrane change procedure and run calibration again.
5) E31 on temp. and D.O. Reading Reason:
a)The inside wire of the probe is broken
Solution: Could not be fixed by user.
Welcome to download the more details from the following linkage,