8930 AZ EB
Sound Level Calibrator
Features
Professional 8930 AZ Sound Level Meter Calibrator features:
* Meets IEC 60942:2018, type 2 class standard
* Stay accurate even in extreme room humidity condition and high altitude places
* Suitable for all sound level meter with microphone head mechanical compliant with IEC 61094-1 or IEC 61094-4, such as 13.2 mm
* Calibration frequency at 1K Hz with three level 114dB, 94dB and 104 dB
* Vivid LED display with stability and low power indicator
* Power up by batteries, easy to carry around
* Compatible with all AZ made sound level meter, includes model 8922, 8921, 8925 and 8928
Application
Application of Professional 8930 AZ Noise Tester Calibrator:
The compact 8930 AZ Sound Level Meter Calibrator is a portable, battery operated sound source, suitable for many portable sound level meters.
When a microphone from a sound level meter is inserted into this calibrator, the sound level meter picks up the signal and displays the signal in dB (decibel) units. The sound level meter should be adjusted (if necessary) to match the calibrator's output signal as closely as possible.
The generated sound pressure level is equal to 94dB, 104dB and 114 dB at 1000 Hz frequency. The 1000 Hz frequency allows calibrating sound level meters with any frequency weighting (LIN, A, B, ...), without applying any correction factor.
It can be conveniently used both in laboratory and in the field. The 114 dB sound level allows performing calibrations even in high background noise environments.
The Digital 8930 AZ Sound Level Meter Calibrator was being designed and manufactured by AZ Instrument, a well experienced manufacturer for Environmental Measuring Instrument.
Specification
Specification of 8930 AZ Sound Level Calibrator:
Model | 8930 | ||
Calibration Frequency | 1000 Hz | ||
Frequency Accuracy | ±1.7% | ||
Sound Pressure Level | 94 dB, 104 dB, 114 dB | ||
Sound Pressure Accuracy | ±0.4 dB | ||
Stabilization Time | 10 Seconds | ||
Total Distortion | <3% | ||
Ambient Condition Influence: Temperature and Humidity Influence Static Pressure Influence |
<0.4 dB at 0~40℃, 25~90% RH <0.1 dB at 65~108 kPa |
||
Stability Levels Short-term Stability (<60s) Stability after 1 year with normal use |
±0.15 dB ±0.35 dB |
||
Battery Life Time | 50 hours typical | ||
LED Indicator | 1 Green/Red: ready/low power 3 Green: 114/104/64 dB |
||
Operating Temperature | 20~30℃ | ||
Operating RH% | 30~60% RH | ||
Storage Temperature | -20~50℃ | ||
Storage RH% | Humidity <90% | ||
Meter Size | 60 (L) * 130(W) * 37.5 (H) mm | ||
Microphone Head Cavity (dia. mm) |
|
||
Weight | ~400g | ||
Battery | AAA Alkaline Batteries * 2 PCS | ||
Standard Package | Meter / Battery / Manual / Paper Box | ||
Picture |