88185 AZ EB

Single Use Dry Ice Ultra Low Temp. PDF Recorder

Welcome to check the Features, Application, Specification, YouTube of 88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Recorder as follows
Features

AZ 88185 Single-Use Ultra-Low Temp PDF Recorder is developed under high-quality manufacturing processes to ensure reliability and safety:

EN 12830: Tested in accordance with EN 12830 for temperature recording devices used in the transport, storage, and distribution of temperature-sensitive goods.

 

RTCA DO-160G: Tested in accordance with RTCA DO-160G environmental conditions and test procedures for airborne equipment.

 

ISO 9001 Facility: Manufactured in an ISO 9001 certified environment.

 

CE EN 61326-1: Designed and tested to meet European EMC safety standards.

 

RoHS & REACH Compliant: We provide a Manufacturer’s Self Declaration, ensuring the product meets environmental safety standards for hazardous substances.

88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Data Logger features:
* Plug & Play. PC software and driver are not required
* Built in PT1000 sensor for dry ice temperature measuring
* Automatically generate PDF report. Report contains statistics summary, diagram and detailed data
* Completely programmable by sender: 6 languages, sampling interval, start delay, alarm delay, unit, alarm range, alarm type, password and user name
* Single trip use to minimize cross-contamination

Application
The 88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Data Logger has a built-in platinum RTD PT1000 temperature sensor that can be used to measure very low temperatures in dry ice environments.

When the recording work is completed, the temperature record can be automatically generated into a PDF or CSV format files, with the security of the temperature record information, confidentiality and not easily deleted. This will meet the requirements of the cold chain supply industry (orchids, flowers, seafood, processed food, perishable food, vaccines and pharmaceuticals), the health and pharmaceutical industry, hospitals or laboratories.

The reasons why different businesses have chosen the 88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Data Logger are listed below,

1. Medical and biotechnology fields:
88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Data Logger can be used to monitor and record the temperature changes of low-temperature refrigerants such as dry ice or liquid nitrogen. This is very important for preserving and transporting some sensitive substances, such as blood, organs, vaccines, etc. 88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Data Logger can automatically generate PDF reports, which are convenient for users to view and analyze data. 88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Data Logger can also set different alarm types and ranges to remind users of temperature anomalies in time. 88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Data Logger can also be used once to reduce the risk of cross-contamination.

2. Food and agriculture fields.
88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Data Logger  can be used to monitor and record the temperature changes of refrigerated or frozen food. This is very important for ensuring the freshness and safety of food. 

3. Industrial and chemical fields.
88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Data Logger can be used to monitor and record the temperature changes of some low-temperature or high-temperature processes or reactions. This is very important for controlling and optimizing the production process. 

The 88165 Dry Ice Temperature USB Data logger was being designed and manufactured by AZ Instrument, a well experienced manufacturer for Environmental Measuring Instrument.
Specification

Specification of 88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Data Logger:

Model 88185
Temperature Range PT1000 RTD Sensor,
-80.0~70.0℃, -112.0~158.0℉
Temperature Resolution 0.1℃, 0.1℉
Temperature Accuracy ±0.5℃ for full range
Logging Type Single Trip Use
Meter Size 104(L) * 28.8(W) * 10.5(H) mm
Sampling Points 8192 readings
On Shelf Time Best within one year time
Operating Temperature -80~70℃ (Logging Status); Room Temp. (PC Mode)
Operating RH% Humidity <80%
Storage Temperature -40~85℃
Storage RH% Humidity <90%
Weight ~20g
Battery 1PC 3.6V Lithium Battery (installed before shipment)
Sampling Interval 30 seconds, 5, 30, 45, 60, 90, 120 minutes
Start Delay 0, 5, 30, 45, 60, 90, 120 minutes
Alarm Range -20~-10℃, -10~0℃, 2~8℃, 0~15℃, 15~25 and others
Alarm Delay 0, 5, 30, 45, 60, 90, 120 minutes
Alarm Type Single Event, Cumulative, Disable
Operation Keys 2 Keys (Start/Stop Key & Mark Key)
LED Indicator REC, Low Battery, High /Low Alarm
Standard Package 1 PC Logger in One Box, Manual, Paper Box
Youtube

Welcome to learn more about the 88185 AZ Single Use Dry Ice Ultra Low Temp. PDF Data Logger from the following product introduction video via YouTube.


AZ 88185 and AZ 88165 dry ice temperature datalogger

AZ 88185 dry ice temperature datalogger Unboxing Video
Other 3

FAQ    AZ 88185 Single-Use Ultra-Low Temp PDF Recorder

Q1: Is AZ 88185 compliant with FDA 21 CFR Part 11?

A1: AZ 88185 does not require any proprietary software to operate or generate reports. Since no software is used throughout the configuration, recording, or data retrieval process, FDA 21 CFR Part 11 software validation is not applicable to this product.

 

Q2: Why is user-configurable important for cold chain monitoring?

A2: Different products require different monitoring rules. User-configurable settings ensure AZ 88181 can meet specific compliance, transport duration, and alarm requirements for each shipment.

 

Q3: Why is AZ 88185 recommended to be used within one year?

A3: AZ 88185 uses an internal battery that naturally self-discharges over time. If the logger is stored for an extended period, the battery voltage may decrease. When such a logger is later used in very low temperatures (e.g. –80 °C), the battery voltage can drop suddenly, which may cause the device to stop recording. For reliable performance, we recommend using AZ 88185 within one year of delivery